<i>Test Results for Chamber M3R4-125</i></font>

Test Results for Chamber M3R4-125


<font color="#00008B" size="5"><i><b>Bar Codes Table. M3R4-125</b></i></font>

Bar Codes Table. M3R4-125

Chamber
M3R4
Chamber
Bar Code
FEBs
Bar Codes
Date Dressing Comments
125 4UAPNB03400125 4UE20900112909
4UE20900111322
4UE20900111923
25-JAN-07 Chamber is in the Pit


<font color="DarkBlue" size="5"><i><b>Gas Leak Test. M3R4-125</b></i></font>

Gas Leak Test. M3R4-125

<font color="DarkRed" size="4"><i>Limit: <= 2mb/hour</i></font>

Limit: <= 2mb/hour

Chamber Gas Leak
[mb/hour]
Test Date Comments
125 .1 Relative methode


<font color="DarkBlue" size="5"><i><b>Durk Current Test. M3R4-125</b></i></font>

Durk Current Test. M3R4-125

<font color="DarkRed" size="4"><i>Limit: Dark Current (sum of 4 gaps) <=100nA at HV=2.9 kV<br />Gas mixture Ar(40)/CO2(55)/CF4(5)<br />The table presents the currents (in nA) measured in the HV line.</i></font>

Limit: Dark Current (sum of 4 gaps) <=100nA at HV=2.9 kV
Gas mixture Ar(40)/CO2(55)/CF4(5)
The table presents the currents (in nA) measured in the HV line.

Chamber 2.4 kV 2.5 kV 2.6 kV 2.7 kV 2.8 kV 2.9 kV Test Date Comments
125 0 1 2 3 6 14 03-FEB-06 T=21.5C, H=23%, P=964mbar



<font color="#00008B" size="5"><i><b>Cosmic HV-Test. M3R4-125</b></i></font>

Cosmic HV-Test. M3R4-125

Table presents counts averaged over 24 pads measured during 300 sec in the mode AND (AB&CD).
Scan1 - HV(AB) fixed @ 2550 V
Scan2 - HV(CD) fixed @ 2550 V


Chamber Scan 2.35 kV 2.45 kV 2.55 kV 2.65 kV 2.75 kV 2.85 kV Date Plot Comments
125 1
2
269
256
331
302
329
329
367
360
478
457
794
657
01-FEB-07


Fig.1 Counts/300s averaged over 24 pads.
AB&CD.
Scan1: HV(AB)=2550V
Scan2: HV(CD)=2550V
Fig.2 Counts distribution in 24 pads.
AB&CD.
HV(AB)=HV(CD)= 2550 V
Fig.3 Counts/300s averaged over 8 pads
in FEB1, FEB2, FEB3.
AB&CD
Scan1: HV(AB)=2550V
Scan2: HV(CD)=2550V


<font color="DarkRed" size="6"><i><b>Cosmic Rays and Noise Counts from Internal Counters. M3R4-125</b></i></font>

Cosmic Rays and Noise Counts from Internal Counters. M3R4-125

Table presents counting rates (counts per 100 sec) averaged over 24 pads in bi-gap AB or in bi-gap CD measured with Thr=16fC in all channels applying HV=2550V at one of the gaps (A,B,C,D) while the other gaps are at HV=0. Also presented are the noise rates (counts per 1 sec) averaged over 48 pads (in bi-gap AB and in bi-gap CD) measured at HV=0 in all gaps with Thr=8.5 fC and with Thr=12 fC in all channels. The thresholds were set individually for each channel according to the relation:
Thr(reg.units)= Thr(fC)*S(reg.un/fC)-OFFSet(reg.un)+Thr_Max(reg.un). Here,
Thr_Max is the threshold at Maximum of noise rate;
OFFSet=17 reg.units (corresponding to 40mV);
S=3.40 reg.units/fC is the charge sensitivity (corresponding to Sens=7.5mv/fC accepted for M3R4 chambers);


HV(ON) HV(OFF)
Chamber HV(A)=2550
GapAB
HV(B)=2550
GapAB
HV(C)=2550
GapCD
HV(D)=2550
GapCD
Thr=8.5 fC Thr=12 fC Atm.Pres. Date Plots Status
125 460.7 445.5 446.7 439.2 9804.3 5.4 Pa=974.4 mb 01-FEB-07

Fig.1 Counts distribution over pads in bi-gap AB (odd channels) and in bi-gap CD (even channels) at various HV combinations:
Left_Up: HV(A)=2550V; HV(B,C,D)=0;
Right_Up: HV(B)=2550V; HV(A,C,D)=0;
Left_Down: HV(C)=2550V; HV(A,B,D)=0;
Right_Down: HV(D)=2550V; HV(A,B,C)=0;
Thr=16fC in all channels. Measurements time 100 sec.
Fig.2 Counts distribution over pads in bi-gap AB (odd channels) and bi-gap CD (even channels) at HV=0 in all gaps
Thr=8.5 fC (upper plot) and in Thr=12 fC (down plot) in all channels. Measurements time 1 sec.

<font color="DarkRed" size="6"><i><b>Thresholds Scan. M3R4-125</b></i></font>

Thresholds Scan. M3R4-125

Tables below present:
- Thr_Max(r.u.), Thresholds in register units at Maximum of noise rate;
- N_Max, Noise rate at Thr_Max in counts/sec;
- D_100 = Thr_100(r.u.) - Thr_Max(r.u.) where Thr_100(r.u.) is the threshold in register units at noise rate = 100/s.
The following expression relates the real threshold (measured from the base-line of physical signals) with the thresholds determined by the register units:
Thr(fC)*S(r.u./fC) = Thr(r.u.) - Thr_Max(r.u.)+ OFFSet(r.u.).
Here, OFFSet=17 register units;
S=3.40 r.u./fC is the charge sensitivity corresponding to S=8mV/fC taken for M3R4 chambers;
(Note, that 1 register unit = 2.35 mV)

FEB-1. Mean values. FEB-2. Mean values. FEB-3. Mean values.
Chamber Thr_Max
[r.u.]
N_Max
[counts/s]
D_100
[r.u.]
Thr_Max
[r.u.]
N_Max
[counts/s]
D_100
[r.u.]
Thr_Max
[r.u.]
N_Max
[counts/s]
D_100
[r.u.]
Test Date Plots Comments
125 75 587852 13 80 679502 14 75 661134 14 01-FEB-07 Sens=7.5mV/fc Th=15fc
<font color="DarkRed" size="5"><i><b>Thr_Max in register units for M3R4-125</b></i><br /></font>

Thr_Max in register units for M3R4-125

FEB
Channels
Ch_1 Ch_2 Ch_3 Ch_4 Ch_5 Ch_6 Ch_7 Ch_8 Ch_9 Ch_10 Ch_11 Ch_12 Ch_13 Ch_14 Ch_15 Ch_16
FEB-1 73 81 80 78 67 82 74 69 72 80 73 77 68 76 70 80
FEB-2 80 85 83 83 72 79 77 83 80 84 85 81 76 85 76 73
FEB-3 73 81 68 79 74 75 73 74 71 79 78 76 80 71 70 76
<font color="DarkRed" size="5"><i><b>Noise rate at Thr_Max [counts/sec] for M3R4-125</b></i><br /></font>

Noise rate at Thr_Max [counts/sec] for M3R4-125

FEB
Channels
Ch_1 Ch_2 Ch_3 Ch_4 Ch_5 Ch_6 Ch_7 Ch_8 Ch_9 Ch_10 Ch_11 Ch_12 Ch_13 Ch_14 Ch_15 Ch_16
FEB-1 462902 839717 629370 670757 312619 722110 472176 479541 507803 423265 613067 638621 505733 608176 806811 712964
FEB-2 566134 760623 809530 728362 591671 562840 537404 470904 708389 546814 683869 584694 702077 628931 1115052 874732
FEB-3 563959 625208 583033 717981 779874 659462 419672 481287 999045 367096 942598 689260 634687 653722 778168 683086
<font color="DarkRed" size="5"><i><b><FONT FACE="Symbol">D_100 </FONT> - in register units for M3R4-125</b></i><br /></font>

D_100 - in register units for M3R4-125

FEB
Channels
Ch_1 Ch_2 Ch_3 Ch_4 Ch_5 Ch_6 Ch_7 Ch_8 Ch_9 Ch_10 Ch_11 Ch_12 Ch_13 Ch_14 Ch_15 Ch_16
FEB-1 12 15 13 13 11 14 13 12 12 12 12 14 13 12 15 14
FEB-2 13 13 15 15 14 13 13 13 13 14 15 13 13 14 16 14
FEB-3 14 14 13 14 14 13 14 13 15 13 15 14 15 15 15 16
<font color="DarkRed" size="5"><i><b>Noise counts vs Thresholds. M3R4-125</b></i></font>

Noise counts vs Thresholds. M3R4-125


Noise rates vs Thresholds for FEB-1 Noise rates vs Thresholds for FEB-2 Noise rates vs Thresholds for FEB-3
<font color="DarkRed" size="5"><i><b>PLOTS of Thresholds Scan Results. M3R4-125</b></i></font>

PLOTS of Thresholds Scan Results. M3R4-125

Scan results for FEB-1 (left plots), for FEB-2 (middle plots), for FEB-3 (right plots).
Thr_Max vs channel number (upper plot)
N_Max vs channel number (middle plot)
D_100 vs channel number (lower plot)



For comments, contact: Petr.Neustroev@cern.ch