<i>Test Results for Chamber M3R4-32</i></font>

Test Results for Chamber M3R4-32


<font color="#00008B" size="5"><i><b>Bar Codes Table. M3R4-32</b></i></font>

Bar Codes Table. M3R4-32

Chamber
M3R4
Chamber
Bar Code
FEBs
Bar Codes
Date Dressing Comments
32 4UAPNB03400032 4UE20900114677
4UE20900111249
4UE20900113571
20-JUN-07 Chamber is in the Pit


<font color="DarkBlue" size="5"><i><b>Gas Leak Test. M3R4-32</b></i></font>

Gas Leak Test. M3R4-32

<font color="DarkRed" size="4"><i>Limit: <= 2mb/hour</i></font>

Limit: <= 2mb/hour

Chamber Gas Leak
[mb/hour]
Test Date Comments
32 .1 Relative methode


<font color="DarkBlue" size="5"><i><b>Durk Current Test. M3R4-32</b></i></font>

Durk Current Test. M3R4-32

<font color="DarkRed" size="4"><i>Limit: Dark Current (sum of 4 gaps) <=100nA at HV=2.9 kV<br />Gas mixture Ar(40)/CO2(55)/CF4(5)<br />The table presents the currents (in nA) measured in the HV line.</i></font>

Limit: Dark Current (sum of 4 gaps) <=100nA at HV=2.9 kV
Gas mixture Ar(40)/CO2(55)/CF4(5)
The table presents the currents (in nA) measured in the HV line.

Chamber 2.4 kV 2.5 kV 2.6 kV 2.7 kV 2.8 kV 2.9 kV Test Date Comments
32 10 14 16 18 20 24 11-OCT-05 chamber is OK



<font color="#00008B" size="5"><i><b>Cosmic HV-Test. M3R4-32</b></i></font>

Cosmic HV-Test. M3R4-32

Table presents counts averaged over 24 pads measured during 300 sec in the mode AND (AB&CD).
Scan1 - HV(AB) fixed @ 2550 V
Scan2 - HV(CD) fixed @ 2550 V


Chamber Scan 2.35 kV 2.45 kV 2.55 kV 2.65 kV 2.75 kV 2.85 kV Date Plot Comments
32 1
2
298
291
339
337
335
335
371
375
466
496
866
890
25-JUN-07


Fig.1 Counts/300s averaged over 24 pads.
AB&CD.
Scan1: HV(AB)=2550V
Scan2: HV(CD)=2550V
Fig.2 Counts distribution in 24 pads.
AB&CD.
HV(AB)=HV(CD)= 2550 V
Fig.3 Counts/300s averaged over 8 pads
in FEB1, FEB2, FEB3.
AB&CD
Scan1: HV(AB)=2550V
Scan2: HV(CD)=2550V


<font color="DarkRed" size="6"><i><b>Cosmic Rays and Noise Counts from Internal Counters. M3R4-32</b></i></font>

Cosmic Rays and Noise Counts from Internal Counters. M3R4-32

Table presents counting rates (counts per 100 sec) averaged over 24 pads in bi-gap AB or in bi-gap CD measured with Thr=16fC in all channels applying HV=2550V at one of the gaps (A,B,C,D) while the other gaps are at HV=0. Also presented are the noise rates (counts per 1 sec) averaged over 48 pads (in bi-gap AB and in bi-gap CD) measured at HV=0 in all gaps with Thr=8.5fC and with Thr=10fC in all channels. The thresholds were set individually for each channel according to the relation:
Thr(reg.units)= Thr(fC)*S(reg.un/fC)-OFFSet(reg.un)+Thr_Max(reg.un). Here,
Thr_Max is the threshold at Maximum of noise rate;
OFFSet=17 reg.units (corresponding to 40mV);
S=3.40 reg.units/fC is the charge sensitivity (corresponding to Sens=7.5mv/fC accepted for M3R4 chambers);


HV(ON) HV(OFF)
Chamber HV(A)=2550
GapAB
HV(B)=2550
GapAB
HV(C)=2550
GapCD
HV(D)=2550
GapCD
Thr=8.5fC Thr=10fC Atm.Pres. Date Plots Status
32 475.3 454.8 465.4 450.5 3239.5 155.9 Pa= 961 mb 22-JUN-07

Fig.1 Counts distribution over pads in bi-gap AB (odd channels) and in bi-gap CD (even channels) at various HV combinations:
Left_Up: HV(A)=2550V; HV(B,C,D)=0;
Right_Up: HV(B)=2550V; HV(A,C,D)=0;
Left_Down: HV(C)=2550V; HV(A,B,D)=0;
Right_Down: HV(D)=2550V; HV(A,B,C)=0;
Thr=16fC in all channels. Measurements time 100 sec.
Fig.2 Counts distribution over pads in bi-gap AB (odd channels) and bi-gap CD (even channels) at HV=0 in all gaps
Thr=8.5fC (upper plot) and in Thr=10fC (down plot) in all channels. Measurements time 1 sec.

<font color="DarkRed" size="6"><i><b>Thresholds Scan. M3R4-32</b></i></font>

Thresholds Scan. M3R4-32

Tables below present:
- Thr_Max(r.u.), Thresholds in register units at Maximum of noise rate;
- N_Max, Noise rate at Thr_Max in counts/sec;
- D_100 = Thr_100(r.u.) - Thr_Max(r.u.) where Thr_100(r.u.) is the threshold in register units at noise rate = 100/s.
The following expression relates the real threshold (measured from the base-line of physical signals) with the thresholds determined by the register units:
Thr(fC)*S(r.u./fC) = Thr(r.u.) - Thr_Max(r.u.)+ OFFSet(r.u.).
Here, OFFSet=17 register units;
S=3.40 r.u./fC is the charge sensitivity corresponding to S=8mV/fC taken for M3R4 chambers;
(Note, that 1 register unit = 2.35 mV)

FEB-1. Mean values. FEB-2. Mean values. FEB-3. Mean values.
Chamber Thr_Max
[r.u.]
N_Max
[counts/s]
D_100
[r.u.]
Thr_Max
[r.u.]
N_Max
[counts/s]
D_100
[r.u.]
Thr_Max
[r.u.]
N_Max
[counts/s]
D_100
[r.u.]
Test Date Plots Comments
32 78 465360 13 75 454460 12 77 476736 13 22-JUN-07
<font color="DarkRed" size="5"><i><b>Thr_Max in register units for M3R4-32</b></i><br /></font>

Thr_Max in register units for M3R4-32

FEB
Channels
Ch_1 Ch_2 Ch_3 Ch_4 Ch_5 Ch_6 Ch_7 Ch_8 Ch_9 Ch_10 Ch_11 Ch_12 Ch_13 Ch_14 Ch_15 Ch_16
FEB-1 75 82 76 79 74 81 81 73 83 72 81 78 77 78 73 82
FEB-2 77 82 71 82 73 76 80 83 69 76 67 71 64 78 66 80
FEB-3 76 72 83 81 80 76 73 86 73 74 71 79 75 80 82 73
<font color="DarkRed" size="5"><i><b>Noise rate at Thr_Max [counts/sec] for M3R4-32</b></i><br /></font>

Noise rate at Thr_Max [counts/sec] for M3R4-32

FEB
Channels
Ch_1 Ch_2 Ch_3 Ch_4 Ch_5 Ch_6 Ch_7 Ch_8 Ch_9 Ch_10 Ch_11 Ch_12 Ch_13 Ch_14 Ch_15 Ch_16
FEB-1 470200 686520 419134 489032 334110 444840 423380 410010 375385 365467 445534 470013 424144 667635 520236 500119
FEB-2 718705 503000 566491 392101 278533 444489 394627 412922 561671 358216 359496 342347 374947 413340 620230 530247
FEB-3 502508 618149 562746 412931 361803 371960 335362 502741 514732 427218 514033 348956 511144 538209 619243 486037
<font color="DarkRed" size="5"><i><b><FONT FACE="Symbol">D_100 </FONT> - in register units for M3R4-32</b></i><br /></font>

D_100 - in register units for M3R4-32

FEB
Channels
Ch_1 Ch_2 Ch_3 Ch_4 Ch_5 Ch_6 Ch_7 Ch_8 Ch_9 Ch_10 Ch_11 Ch_12 Ch_13 Ch_14 Ch_15 Ch_16
FEB-1 13 15 12 14 13 13 12 13 12 13 12 14 13 15 14 14
FEB-2 13 14 13 12 11 12 12 12 12 12 12 11 12 12 14 14
FEB-3 13 14 13 12 13 13 11 12 11 12 13 12 13 13 14 14
<font color="DarkRed" size="5"><i><b>Noise counts vs Thresholds. M3R4-32</b></i></font>

Noise counts vs Thresholds. M3R4-32


Noise rates vs Thresholds for FEB-1 Noise rates vs Thresholds for FEB-2 Noise rates vs Thresholds for FEB-3
<font color="DarkRed" size="5"><i><b>PLOTS of Thresholds Scan Results. M3R4-32</b></i></font>

PLOTS of Thresholds Scan Results. M3R4-32

Scan results for FEB-1 (left plots), for FEB-2 (middle plots), for FEB-3 (right plots).
Thr_Max vs channel number (upper plot)
N_Max vs channel number (middle plot)
D_100 vs channel number (lower plot)



For comments, contact: Petr.Neustroev@cern.ch